Abstract

  • Background: Noninvasive monitoring would likely improve trauma care. Using laser technology, we monitored the oxygen saturation in retinal vessels during exsanguination and hypoxia. Methods: Seven anesthetized swine were bled at 0.4 mL/kg/min for 40 minutes. During exsanguination, retinal venous saturation (S(rv)O2) was measured using an eye oximeter, and central venous saturation (S(v)O2) was measured using a fiber-optic catheter. After the shed blood was reinfused, the FiO2 was progressively decreased from 0.97 to 0.07. Femoral artery oxygen saturation (S(a)O2) and retinal artery oxygen saturation (S(ra)O2) were measured at each increment. Results: During exsanguination, S(rv)O2 correlated with blood loss (r = -0.93) and S(v)O2 (r = 0.94). S(ra)O2 correlated with S(a)O2 during incremental hypoxia (R2 = 0.93 ± 0.15). Conclusions: In this model of exsanguination, retinal venous oxygen saturation correlates with blood volume and with central venous oxygen saturation. The S(ra)o2 correlates with S(a)O2 during graded hypoxia. Use of an eye oximeter to noninvasively monitor trauma patients appears promising and warrants further study.
  • Authors

    Digital Object Identifier (doi)

    Author List

  • Denninghoff KR; Smith MH; Chipman RA; Hillman LW; Jester PM; Hughes CE; Kuhn F; Rue LW
  • Start Page

  • 29
  • End Page

  • 34
  • Volume

  • 43
  • Issue

  • 1