A multicenter evaluation of lipid profiling with a compact analyzer (Miles Clinistat).

Academic Article

Abstract

  • We evaluated the Clinistat Analyzer (Miles Inc., Diagnostics Division, Elkhart, IN) for measuring cholesterol, triglycerides, and high-density lipoprotein (HDL) cholesterol at three medical centers. The system, based on multilayer film technology, uses precalibrated, dry film reagent disks. Ten microliters of serum is applied to the dry film reagent disk in the test procedure. For HDL-cholesterol measurement, serum is pretreated by precipitation with phosphotungstic acid and magnesium chloride. Total precision (CVs) of each of the three assays was less than or equal to 5%. The assay ranges were linear and satisfactory for clinical use. Patients' results compared well with established methods. No significant interferences were found with hemolysis, icterus, and lipemia.
  • Keywords

  • Chemistry, Clinical, Cholesterol, Cholesterol, HDL, Evaluation Studies as Topic, Humans, Quality Control, Triglycerides
  • Author List

  • Hardy RW; Ng RH; Hill RE; Walker S; Sparks KM; Kent S; Lakes C; Hiar CE; Valdes R; Statland BE
  • Start Page

  • 1437
  • End Page

  • 1439
  • Volume

  • 38
  • Issue

  • 8 Pt 1