Thickness of molybdenum filter and squared contrast-to-noise ratio per dose for digital mammography

Academic Article

Abstract

  • OBJECTIVE. The objective of our study was to test whether the lesion-tissue contrast-to-noise ratio (CNR) at a given dose level can be improved by increasing the thickness of the molybdenum (Mo) filter currently used in digital mammography. MATERIALS AND METHODS. We studied how the CNR between breast and a 5-mm simulated infiltrating ductal carcinoma (IDC) embedded in a 5-cm-thick breast changes with Mo filter thickness. We performed phantom imaging experiments by modifying the filter wheel of a Senographe 2000D unit with Mo filters that ranged from 15 to 90 μm thickness. A 5-cm-thick 50% glandular-50% adipose breast phantom with a 5-mm insert simulating IDC was used as the phantom for all the cases. The CNRs between the breast phantom and the IDC insert were measured, and average glandular doses were calculated using a filtration-dependent X-ray spectra model and a breast dosimetry model based on a validated Monte Carlo simulation. RESULTS. The lesion-tissue CNR at a given dose level increases with increasing Mo filter thickness from 15 to 90 μm. The measured squared CNR per dose increased by 8%, 14%, 17%, and 17% for 45-, 60-, 75-, and 90-μm Mo filters, respectively, compared with the standard 30-μm Mo filter. Meanwhile, the exposure times were increased by 35% (45 μm), 71% (60 μm), 177% (75 μm), and 229% (90 μm). CONCLUSION. Increasing Mo filter thickness from 30 to 60 μm can increase lesion-tissue squared CNR per dose by 14% with a tolerable increase in the duration of exposure. © American Roentgen Ray Society.
  • Authors

    Digital Object Identifier (doi)

    Author List

  • Nishino TK; Wu X; Johnson RF
  • Start Page

  • 960
  • End Page

  • 963
  • Volume

  • 185
  • Issue

  • 4