Reproducibility of operator processing for radiation dosimetry

Academic Article

Abstract

  • Reproducibility of operator processing for radiation dose and biological half-life was assessed for radioimmunotherapy. Mean coefficient of variation for intra-operator consecutive processing and for inter-operator processing was less than 15% for all tissues. The mean coefficient of variation for intraoperator processing over 2 wk or inter-operator processing comparing an experienced and less experienced operator was generally greater, and particularly so for tumors. Satisfactory reproducibility was achievable using visual determination of regions of interests after 80 h of training.
  • Authors

    Published In

    Digital Object Identifier (doi)

    Author List

  • Shen S; DeNardo GL; DeNardo SJ; Yuan A; DeNardo DA; Lamborn KR
  • Start Page

  • 77
  • End Page

  • 83
  • Volume

  • 24
  • Issue

  • 1