Calibration setting numbers for dose calibrators for the PET isotopes 52Mn, 64Cu, 76Br, 86Y, 89Zr, 124I

Academic Article

Abstract

  • © 2016 Elsevier Ltd. For PET radionuclides, the radioactivity of a sample can be conveniently measured by a dose calibrator. These devices depend on a "calibration setting number", but many recommended settings from manuals were interpolated based on standard sources of other radionuclide(s). We conducted HPGe gamma-ray spectroscopy, resulting in a reference for determining settings in two types of vessels containing one of several PET radionuclides. Our results reiterate the notion that in-house, experimental calibrations are recommended for different radionuclides and vessels.
  • Digital Object Identifier (doi)

    Author List

  • Wooten AL; Lewis BC; Szatkowski DJ; Sultan DH; Abdin KI; Voller TF; Liu Y; Lapi SE
  • Start Page

  • 89
  • End Page

  • 95
  • Volume

  • 113